BIO / Medical
isolator
Kiyon`s BIO Glovebox protects operators in the
pharma industries, the air filtration system provides
customized design and configuration.
Technical Specification
2-4-2 |
GloveFAST Cyto Pharma 2-5-4 |
2-6-4 |
||
---|---|---|---|---|
Isolator Dimensions |
Outer dim. (WxDxH) complete | 2510 x 870 x 2300 mm | 2815 x 870 x 2300 mm | 3120 x 870 x 2300 mm |
Inner dim. (WxDxH) work chamber | 1192 x 580 x 740 mm | 1497 x 580 x 740 mm | 1802 x 580 x 740 mm | |
Gloves material | Neoprene | |||
Sleeves material | CSM | |||
Sleeve dimensions | Dimensions at wrist 90 mm-Glove port Ø 300/250 mm | |||
Electrical Connection & Weight |
Weight (approx. kg) | 550 | 580 | 620 |
Lighting level (Lux) | ≥ 2000 | |||
Noise level (dB(A)) | ≤ 53 | ≤ 55 | ≤ 58 | |
Working Parameters | Operating pressure | -75 Pa | ||
Downflow air speed | 0.45 m/s | |||
Exhaust Flow rate | 300 m3/h |
Top Quality
- Air quality in grade A (unidirectional Laminar Flow) as per EU-GMP Guide within the entire work zone
- Easy cleaning of inner and outer surface
- Automatic pressure decay test (via optional ATV valve) with results of the test stored in the isolator back up memory.
- Negative operating pressure (-75Pa) with laminar airflow and actively regulated blowers
- Aseptic glove-changing system
- Easy monitoring of all parameters, test and maintenance
- Ergonomically designed for easy operation, good visibility and high operators' comfort
- High air-change rate within the work zone and the transfer hatches for rapid clean-up of contaminated air
- Foot switch controlled inner doors
Compliance
- Filtration: EN 1822 / IEST-RP-CC00 1.3
- Air Cleanliness: ISO 14644-1
- Containment: ISO 14644-7
- Electrical Safety: EN 61010-1 / IEC 61010-1
- Safety Cabinets, Glove Boxes for Cytotoxic Substances and Other CMR drugs : DIN 12980:2017